Medical Device Classification under US FDA

Next: FDA Quality Management System Regulation (QMSR) →

Your device's classification decides a lot: which controls apply, and which route you take to the U.S. market. FDA has grouped about 1,700 generic types of devices into 16 medical specialties, called panels. Each type is assigned to one of three classes, based on how much control is needed to keep the device safe and effective.

The three classes

The classes build on each other. Every device, whatever its class, has to meet general controls. Class II adds special controls on top. Class III devices need premarket approval as well.

Increasing risk and regulatory control →
510(k) unless exempt
General controls
Class ILowest risk
510(k) unless exempt
Special controls
General controls
Class IIModerate risk
Usually a PMA
Premarket approval (PMA)
General controls
Class IIIHighest risk

What decides the class

FDA looks at what the device is for and how much harm it could do. The first comes down to intended use and indications for use.

  • Intended use is the general purpose of the device or its function, and it encompasses the indications for use.
  • Indications for use describes the disease or condition the device will diagnose, treat, prevent, cure or mitigate, including a description of the patient population the device is intended for.

For example, a light-cure dental composite, the intended use might be "restoration of teeth." The indications for use might be "direct restorations of Class I to V cavities in anterior and posterior teeth, and core build-up."

The second factor considered is the risk. The more a device could harm a patient or user, the higher its class. Class I holds the lowest-risk devices and Class III the highest.

What it means for market entry

Your device class decides which premarket submission FDA expects before you can sell. If your device is Class I or II and is not exempt, you need a 510(k). Exempt devices skip the 510(k), but the limitations on exemptions still apply.

Class III devices need a premarket approval application (PMA), with one exception. A preamendments device, meaning one that was on the market before the 1976 Medical Device Amendments or is substantially equivalent to one, goes through a 510(k) as long as FDA has not called for PMAs for that type.

Finding your device's class

To find your device's class, first find the regulation that classifies it. You can get there from the classification database or from the panel list.

  • Search the database: Enter part of your device name in FDA's Product Classification Database. In most cases it shows the classification regulation.
  • Or start from the panel: If you know your device's medical specialty panel, open it in the Code of Federal Regulations (CFR). Each panel begins with a list of the devices classified in it.
  • Read the regulation: It describes the device, states its class, and tells you about any exemptions.
  • Note the product code: This three-letter code is also the classification number used on the Medical Device Listing form

Still unsure? Ask FDA

When you need a formal answer on how FDA classifies your device, you can file a 513(g) request. There is a user fee, though eligible small businesses can pay a reduced one through FDA's Small Business Determination program.

Not sure where your device falls?

Share your device description and intended use, and we'll help you find the right classification and the route that goes with it.

Last reviewed: October 2026