If your device goes through the 510(k) Program, you get to choose how you submit. There are three routes: Traditional, Special and Abbreviated. Each one suits a different kind of submission, and the right pick can save you time.
The Traditional 510(k) is the original submission type, set out in 21 CFR 807. In 1998, FDA added the Special and Abbreviated programs to make it easier to review certain kinds of 510(k) submissions. You can use either one when your submission meets the right factors.
Choosing one of the alternative programs does not limit what FDA can ask for. The agency can still request any information the law and regulations allow, so it can decide whether your device is substantially equivalent to a predicate device.
| Type | Usually used for | FDA review timeline |
|---|---|---|
| Traditional | Any original 510(k), or a change to a device already cleared through 510(k) | Generally, within 90 days |
| Special | A change to your own existing device, where well-established methods can evaluate it | Generally, within 30 days |
| Abbreviated | A submission that relies on FDA guidance, special controls or consensus standards | Generally, within 90 days |
This is the all-purpose option. You can use it for any original 510(k), and when you are changing a device that was already cleared through 510(k).
After the submission clears FDA's acceptance review (under the Refuse to Accept policy), FDA generally reviews a Traditional 510(k) within 90 days of receiving it.
A Special 510(k) is for manufacturers who are changing their own existing device. It works when there are well-established methods to evaluate the change, and when the results can be reviewed in a short summary or risk analysis format.
When it fits
A design or labeling change to an existing device, including certain changes to the indications for use, may qualify when all of these are true:
- The manufacturer legally authorized to market the existing device is the one submitting the change.
- Performance data are unnecessary, or, if they are necessary, well-established methods exist to evaluate the change.
- All the performance data needed to support substantial equivalence can be reviewed in a summary or risk analysis format.
A case where it usually doesn't fit. Devices that make a biological product at the point of care are generally a poor match. There is probably no well-established way to evaluate those changes, and the performance data may not be reviewable as a summary or risk analysis.
After the acceptance review, FDA generally reviews a Special 510(k) within 30 days of receipt.
Choose an Abbreviated 510(k) when your submission leans on one or more of these:
- FDA guidance documents
- Demonstrated compliance with the special controls for your device type
- Voluntary consensus standards
An Abbreviated 510(k) still has to include every element required for a Traditional 510(k) under 21 CFR 807.87. What changes is how you present the evidence. To make the review easier, you provide summary reports on how you used the guidance documents or special controls, or declarations of conformity to standards FDA recognizes.
If you rely on a voluntary consensus standard in a general way, explain the basis for using it. Include the underlying information or data that shows how you applied the standard.
After the acceptance review, FDA generally reviews an Abbreviated 510(k) within 90 days of receipt.
Which one fits?
A quick way to think about it:
You have an original 510(k), or a change to a cleared device, and none of the shortcuts apply.
You are changing your own device, and well-established methods exist to evaluate the change.
Your evidence rests on FDA guidance, special controls or recognized consensus standards.
Not sure which type suits your device?
Tell us about your device and the change you have in mind. We will help you choose the route that fits and build the submission around it. Contact us today
We support you through:
- Device classification and predicate device identification
- Substantial equivalence comparison and technical rationale
- Compilation of performance testing, biocompatibility, and labeling documentation
- Preparation and formatting of the complete 510(k) submission package
- Responding to FDA queries and Additional Information requests
Our goal is to help you build a submission that is scientifically sound, well-documented, and positioned for a smooth FDA review.